Zone 0 / Class I Div. 1
TO MEASURE
PRESSURE
TECHNICAL DATA
Ambient temperature 0 °C to 50 °C
Storage temperature -40 to 60 °C
Pressure module output LEMO connector
Relative humidity
(% RH operating without condensation)
95 % (+10 °C... +30 °C); 75 % (+30 °C... +40 °C);
45 % (+40 °C... +50 °C); 35 % (+50 °C... +55 °C)
Dimensions 110 x 90 x 45 mm
Weight approx. 300 g
PRESSURE MODULE SPECIFICATIONS
Model Range (approx.) Resolution
Reference
Uncertainty
(+23 ± 3 °C)
High Side
media
Low Side
media
Fitting
material
Max.
over-
pressure
2
Differential
700P01Ex
700P24Ex
25 mbar
1000 mbar
0.01 mbar
0.1 mbar
0.2 %
0.03 %
dry
1
316 SS
dry
dry
316 SS
316 SS
3x
3x
Gauge
700P05Ex
700P06Ex
700P27Ex
700P09Ex
2 bar
7 bar
20 bar
100 bar
0.1 mbar
0.7 mbar
1 mbar
10 mbar
0.03 %
0.03 %
0.03 %
0.03 %
316 SS
316 SS
316 SS
316 SS
N/A
N/A
N/A
N/A
316 SS
316 SS
316 SS
316 SS
3x
3x
3x
2x
Absolute
700PA4Ex 1000 mbar 0.1 mbar 0.05 % 316 SS N/A 316 SS
3x
High
700P29Ex 200 bar 0.01 bar 0.05 % C276 N/A C276
2x
1
“Dry” indicates dry air or non-corrosive gas as compatible media.
“316 SS” indicates media compatible with Type 316 Stainless Steel.
“C276” indicates media compatible with Hastelloy C276.
2
Maximum overpressure specication includes common mode pressure.